ACR Electronics SmartReader Plus Manuel d'utilisateur Page 86

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 103
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 85
SmartReader Data Loggers
78
Copyright © 2010 ACR Systems Inc. All Rights Reserved.
black leads are stripped and tinned.
Figure C-2: Resolution Chart for the ET-004 Probe
ET-081 High Temperature Probe
Description: NTC thermistor and cable potted into the end of a 22mm
(7/8") long by 5mm (3/16") diameter nickel-plated eyelet.
R25 Value: 100k ohms
Equation: 47
Range: 10 to 170°C (50 to 335°F)
Max. Temp.: 190°C (375°F)
Resolution: Refer to Figure C-3
Accuracy: ±0.5°C (0.9°F) from the temperature values listed in Table
C-1 over the range of 10 to 170°C (50 to 335°F).
Cable: 6m (20ft) standard length with twisted-pair Teflon® insulated
Vue de la page 85