SmartReader Data Logger
27
Copyright © 2010 ACR Systems Inc. All Rights Reserved.
as (A). The reading should be close to zero if you are using a gauge sensor, or close to
ambient pressure (AP) if you are using an absolute sensor.
5. Enter a Low calibration value using:
· For absolute sensor: -(A-AP)*100/FS
· For gage sensor: -(A*100/FS)
Where:
A = The Realtime reading (in PSI) observed in previous step.
AP = Ambient pressure.
FS = Full scale rating of the logger.
NOTE: Make sure you enter the Low calibration value before going to the next
step.
7. Connect the input port of the logger to the calibrator, ensuring all connections are
completely airtight.
8. Adjust the calibrator to produce a test pressure equal to the full scale (FS) range of the
logger. Record the pressure reading you see as (B).
9. Enter a Mid calibration value using:
· (FS/B-1)*100
Where:
FS = Full scale range of the logger.
B = The real-time reading
10. Save the calibration changes to the logger.
1.6.4 Specifications
The following specifications apply to the pressure sensors used in the SmartReader 4:
Type: Silicon piezoresistive strain gauge
Max. Pressure: 4 x FS
Media: Noncorrosive gases and dry air
Temp. Range:
Operating: -40 to 70°C (-40 to 158°F)
Compensated: 0 to 70°C (32 to 158°F)
Accuracy: 0.5% FS @ 25°C (77°F)
Thermal: over Compensated Range
5 PSI ±1.5% FS
30 PSI ±0.5% FS
100 PSI ±0.5% FS
Linearity: best fit straight line
5 PSI ±0.25% FS
30 PSI ±0.1% FS
100 PSI ±0.1% FS