SmartReader Data Logger
29
Copyright © 2010 ACR Systems Inc. All Rights Reserved.
1.7
SmartReader 5 Thermocouple Logger
The SmartReader 5 Thermocouple Logger can monitor and record temperatures from type J,
K, T, or S type thermocouples.
Figure 5-1: SmartReader 5
1.7.1 Description
The SmartReader 5 has two external thermocouple temperature channels and one internal
thermistor temperature channel. There are two different temperature ranges:
· Channel 1 (Low) is suited for use over a relatively narrow temperature range. It
provides the best resolution readings.
· Channel 2 (High) is useful over a much wider temperature range thus offering greater
application flexibility. Resolution is coarser using this channel.
For information on temperature ranges and resolutions applicable for the different types of
thermocouples and channels you want to use, refer to the Specifications section.
1.7.2 Setup and Use
The internal thermistor temperature channel for the SmartReader 5 is used to simulate a
reference junction for the thermocouples -- it must always be enabled.
1. Connect a J, K, T, or S type thermocouple to one or both channels. Make sure that you
observe polarity when making connections. To do this, connect a thermocouple's
negative wire (usually red) to the "-" terminal of the channel you want to use. Connect the
other (positive) wire to the "+" terminal. The following table lists the ANSI color code for
the thermocouple wires: