Choose whether you want to use the engineering value or a raw value for the input. For
the pressure transducer example, you would select Channel engineering value as you
would be entering voltages and corresponding pressures.
Enter the order of polynomial equation. You will need to enter at least 1 more input-
output pairs than the order of the equation. For example, if you choose a 3rd order
equation, you will need to enter at least 4 input-output pairs.
Enter the first input value and its matching output.
Enter the second input value and its matching output.
Enter the minimum number of input-output pairs required. If more rows are required, click
. If you have added too many rows, remove them by clicking on
Choose whether you will enter the temperature in degrees Celsius or Fahrenheit.
Enter the low input temperature and its matching resistance.
Enter the mid input temperature and its matching resistance.
Enter the high input temperature and its matching resistance.