The mechanical support of the x-ray tube on Cephalometric
and Panoramic units shall be designed
so that the x-ray
tube housing assembly remains stable during an exposure
after positioning.
units by placing
This should be evaluated on intraoral
the x-ray
tube assembly in the
positions/angles commonly used
cephalometric unit should
clinically.
The
be set up
geometry to evaluate mechanical
stability.
in its clinical
While the x-
ray tube assembly is not stationary on Panoramic units
during exposures; it
should
be
maintain the vertical and
rigidly supported
horizontal slit
to
alignment
discussed in Section III-5-B.
III-10. Shielding Devices
The comments made in Section I-4 also generally apply
to dental installations. Typically a lead apron is placed
on the patient to provide ‘whole body” and
gonadal
shielding.
specifically
Thyroid shields are also effective at
reducing thyroid doses As Low As Reasonably Achievable
(ALARA) for dental patients.
44